Intraocular pressure measurements in paediatric glaucoma: A narrative review on accuracy, tolerability, and ease of use.
Medical Journal of Malaysia 2024 March
INTRODUCTION: Numerous tonometers are available to measure intraocular pressure (IOP) in children with glaucoma. This review aims to discuss IOP measurement techniques and principles and compare the accuracy, tolerability and ease of use of available tonometers in measuring IOP in paediatric glaucoma patients.
MATERIALS AND METHODS: A review of observational studies was conducted to discuss the accuracy, tolerability and ease of use of tonometers in measuring IOP in children with glaucoma.
RESULTS: Goldmann applanation tonometry (GAT) and its portable handheld versions remain the gold standard in measuring IOP. Tono-Pen (Reichert Ophthalmic Instruments, Depew, New York, USA) and rebound tonometer (RBT) both correlate well with GAT. Although both tonometers tend to overestimate IOP, Tono-Pen overestimates more than RBT. Overestimation is more remarkable in higher IOP and corneal pathologies (such as but not limited to scarred cornea and denser corneal opacity). RBT was better tolerated than other tonometers in children and was easier to use in children of all ages.
CONCLUSIONS: RBT is the preferred tonometer for measuring IOP in children with glaucoma, as it is less traumatic, time efficient and does not require fluorescein dye or anaesthesia. However, examiners should use a second tonometer to confirm elevated IOP readings from the RBT.
MATERIALS AND METHODS: A review of observational studies was conducted to discuss the accuracy, tolerability and ease of use of tonometers in measuring IOP in children with glaucoma.
RESULTS: Goldmann applanation tonometry (GAT) and its portable handheld versions remain the gold standard in measuring IOP. Tono-Pen (Reichert Ophthalmic Instruments, Depew, New York, USA) and rebound tonometer (RBT) both correlate well with GAT. Although both tonometers tend to overestimate IOP, Tono-Pen overestimates more than RBT. Overestimation is more remarkable in higher IOP and corneal pathologies (such as but not limited to scarred cornea and denser corneal opacity). RBT was better tolerated than other tonometers in children and was easier to use in children of all ages.
CONCLUSIONS: RBT is the preferred tonometer for measuring IOP in children with glaucoma, as it is less traumatic, time efficient and does not require fluorescein dye or anaesthesia. However, examiners should use a second tonometer to confirm elevated IOP readings from the RBT.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app