Hyperkalemia-Induced T Wave Oversensing in an Implantable Cardioverter Defibrillator.

Curēus 2024 Februrary
A 66-year-old female with end-stage renal disease and heart failure with reduced ejection fraction, status post implantable cardioverter defibrillator (ICD) presented to the emergency department with dizziness and fatigue. An electrocardiogram showed sinus rhythm, complete atrioventricular block, and ventricular paced rhythm at 30 beats per minute (bpm). Device interrogation revealed a programmed VVI mode with a lower rate limit of 40 bpm and evidence of T wave oversensing. Serologic studies were remarkable for hyperkalemia (7.9 mmol/dL). The device was initially reprogrammed to provide a higher pacing rate and symptomatic improvement. Both complete AV block and T wave oversensing resolved after correction of hyperkalemia. This case highlights the need for vigilant monitoring of electrolyte imbalances in ICD patients.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app