Comparison and estimation of photoneutron dose produce between 10 MV flattened and unflattened beam in Elekta Versa HD™ medical linac.

BACKGROUND: In a high-energy medical linear accelerator (linac), if the interaction of photon energy is higher than the neutron binding energy of high atomic material, it emits a neutron field through photonuclear (γ, n) reaction.

AIM: The current study, evaluates the photoneutron dose equivalent (PNDE) produced between the 10 MV flattened and unflattened beams as a function of field sizes in the Elekta Versa HD™ linac.

MATERIALS AND METHODS: The PNDE produced from Versa linac was recorded along the patient plane using the bubble detector personal neutron dosimeter and from the measured PNDE values, the theoretical PNDE values were simulated for various field sizes using nonlinear least-squares curve-fitting as a function of a polynomial. The percentage of deviation (PoD) and Chi-square (χ2) tests were performed between the measured and simulated PNDE values to study the reliability and validity.

RESULTS: The results show that the mean PoD between the measured and simulated PNDE values for respective positions of a field size of FF beam was found to be -1.99% for 0.3×0.3, -4.39% for 5×5, -3.868% for 10×10, 0.590% for 15×15, 9.18% for 20×20, -4.133% for 25×25, and 0.467% for 30×30 cm2. Similarly, the mean PoD between the measured and simulated PNDE values for flattening filter-free (FFF) beam was found to be 1.36% for 0.3×0.3, -1.39% for 5×5, -5.38% for 10×10, 4.41% for 15×15, 3.84% for 20×20, 5.69% for 25×25, and -1.75% for 30×30 cm2. The maximum deviation between the measured and simulated PNDE values lies within the range ± 5%.

CONCLUSIONS: From the study, it is observed that the FFF beam produces lesser neutron contamination than the FF beam.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app