Using cryoprobes of different sizes combined with cone-beam computed tomography-derived augmented fluoroscopy and endobronchial ultrasound to diagnose peripheral pulmonary lesions: a propensity-matched study.

Respiratory Research 2024 Februrary 6
BACKGROUND: Endobronchial ultrasound (EBUS) and cone-beam computed tomography-derived augmented fluoroscopy (CBCT-AF) are utilized for the diagnosis of peripheral pulmonary lesions (PPLs). Combining them with transbronchial cryobiopsy (TBC) can provide sufficient tissue for genetic analysis. However, cryoprobes of different sizes have varying degrees of flexibility, which can affect their ability to access the target bronchus and potentially impact the accuracy. The aim of this study was to compare the diagnostic efficacy of cryoprobes of varying sizes in CBCT-AF and EBUS for the diagnosis of PPLs.

METHODS: Patients who underwent endobronchial ultrasound-guided transbronchial biopsy (EBUS-TBB) and TBC combined with CBCT-AF for PPLs diagnosis between January 2021 and May 2022 were included. Propensity score matching and competing-risks regression were utilized for data analysis. Primary outcome was the diagnostic accuracy of TBC.

RESULTS: A total of 284 patients underwent TBC, with 172 using a 1.7-mm cryoprobe (1.7 group) and 112 using a 1.1-mm cryoprobe (1.1 group). Finally, we included 99 paired patients following propensity score matching. The diagnostic accuracy of TBC was higher in the 1.1 group (80.8% vs. 69.7%, P = 0.050), with a similar rate of complications. Subgroup analysis also revealed that the 1.1 group had better accuracy when PPLs were located in the upper lobe (85.2% vs. 66.1%, P = 0.020), when PPLs were smaller than 20 mm (78.8% vs. 48.8%, P = 0.008), and when intra-procedural CBCT was needed to be used (79.5% vs. 42.3%, P = 0.001). TBC obtained larger specimens than TBB in both groups. There is still a trend of larger sample size obtained in the 1.7 group, but there is no statistically different between our two study groups (40.8 mm2 vs. 22.0 mm2 , P = 0.283).

CONCLUSIONS: The combination of TBC with CBCT-AF and EBUS is effective in diagnosing PPLs, and a thin cryoprobe is preferred when the PPLs located in difficult areas.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app