Effect of Disinfection Method and Testing Methodology on the Performance of a Breath-Enhanced Jet Nebulizer.
Pharmaceutics 2023 December 22
National guidelines for cystic fibrosis recommend cleaning and disinfecting nebulizers after each use. We tested two groups of five reusable breath-enhanced nebulizers after 0, 5, 10, 15, 20, 30, 60, 90, 120, 150, and 180 sterilization (baby bottle sterilizer) or cleaning cycles. The nebulizers were operated for 7 min (6 L/min) after loading albuterol (2.5 mg/3 mL), and they were evaluated with and without breathing simulation after cleaning/sterilization (0-180 and 0-60 cycles, respectively). Over the course of 180 cleaning/sterilization cycles, the mean (SD) solution output was 1.33 mL (0.12 mL)/1.29 mL (0.08 mL); the nebulizer mass remaining in the nebulizer was 61.5% (5.2%)/63% (4%); sputtering time was 4.7 min (0.8 min)/4.8 s (0.6 min); inspiratory filter was 19% (3%)/18.5% (2.4%); expiratory filter was 6.7% (1.1%)/6.7% (0.8%); and difference in drug output calculated using the solution output and nebulizer mass was 6.8% (4%)/5.2% (2.9%). Thermal disinfection with a baby-bottle sterilizer did not alter the performance of a reusable breath-enhanced nebulizer. The nebulizer test performed without breathing simulation underestimated its performance. The calculation of the drug output based on the solution output resulted in its overestimation.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app