Comparison of Evolut-R 34 mm Valve and Smaller Evolut-R Valves in Patients Undergoing Transcatheter Aortic Valve Implantation and Determination of Mild Paravalvular Leak Predictors.
Anatolian Journal of Cardiology 2024 January 16
OBJECTIVE: The main purpose of this study was to evaluate and compare the in-hospital, 1-month and 1-year post-procedure outcomes of patients treated with Evolut-R 34 mm and Evolut-R 23/26/29 mm devices. Additionally, the study aimed to identify factors that could predict the occurrence of ≥ mild paravalvular leaks (PVL).
METHODS: Between April 2015 and May 2022, 269 consecutive patients who underwent transcatheter aortic valve implantation (TAVI) with Evolut-R 34 mm (n = 66, 24.5%) and Evolut-R 23/26/29 mm (n = 203, 75.5%) devices in a single center were retrospectively analyzed.
RESULTS: Patients in the Evolut-R 34 mm group had a lower female sex ratio (16.7% vs. 66.5%, P <.001, respectively), ejection fraction (50.7 ± 10.1% vs. 54.5 ± 9.3%, P =.016, respectively), and mean aortic gradient (7.4 ± 3.3 vs. 9.2 ± 5.0, P =.026, respectively) compared to the Evolut-R 23/26/29 mm group. The groups did not exhibit any statistically significant distinctions with regard to technical success, the need for a permanent pacemaker, occurrences of stroke, major vascular complications, PVL, major adverse cardiovascular and cerebrovascular events, or mortality. Peak velocity was confirmed as a significant predictor of ≥mild PVL in both patient groups in the receiver operating characteristic curve analysis. In logistic regression analysis; In patients with Evolut-R 34 mm valve, pre-TAVI aortic valve peak velocity (odds ratio (OR) = 23.202; P =.019) and calcium volume 800 Hounsfield Units (mm3) (OR = 1.017; P <.001) were independent predictors of ≥mild PVL.
CONCLUSION: The Evolut-R 34 mm valve has shown comparable in-hospital results with smaller valve sizes. Pre-TAVI aortic valve peak velocity and calcium volume predicted ≥ mild PVL in Evolut-R 34 mm patients.
METHODS: Between April 2015 and May 2022, 269 consecutive patients who underwent transcatheter aortic valve implantation (TAVI) with Evolut-R 34 mm (n = 66, 24.5%) and Evolut-R 23/26/29 mm (n = 203, 75.5%) devices in a single center were retrospectively analyzed.
RESULTS: Patients in the Evolut-R 34 mm group had a lower female sex ratio (16.7% vs. 66.5%, P <.001, respectively), ejection fraction (50.7 ± 10.1% vs. 54.5 ± 9.3%, P =.016, respectively), and mean aortic gradient (7.4 ± 3.3 vs. 9.2 ± 5.0, P =.026, respectively) compared to the Evolut-R 23/26/29 mm group. The groups did not exhibit any statistically significant distinctions with regard to technical success, the need for a permanent pacemaker, occurrences of stroke, major vascular complications, PVL, major adverse cardiovascular and cerebrovascular events, or mortality. Peak velocity was confirmed as a significant predictor of ≥mild PVL in both patient groups in the receiver operating characteristic curve analysis. In logistic regression analysis; In patients with Evolut-R 34 mm valve, pre-TAVI aortic valve peak velocity (odds ratio (OR) = 23.202; P =.019) and calcium volume 800 Hounsfield Units (mm3) (OR = 1.017; P <.001) were independent predictors of ≥mild PVL.
CONCLUSION: The Evolut-R 34 mm valve has shown comparable in-hospital results with smaller valve sizes. Pre-TAVI aortic valve peak velocity and calcium volume predicted ≥ mild PVL in Evolut-R 34 mm patients.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app