We have located links that may give you full text access.
Dental unit waterline testing practices: an 11-Year retrospective study.
BMC Oral Health 2023 November 16
OBJECTIVES: This retrospective study examined the dental unit waterline (DUWL) testing practices of Saskatchewan dental clinics over a period of 11 years, with an emphasis on their responses after identification of high microbial levels.
MATERIALS AND METHODS: Dental clinics (n = 137) aseptically collected samples of output water from their air/water syringes, handpieces, and ultrasonic scaler lines using Sigma-Aldrich® waterline test kits and delivered them to a quality assurance laboratory. Tests were incubated for seven days at room temperature, and those with heterotrophic plate counts > 500 CFU/mL were reported as failures. Statistical analyses were performed on a database containing 4,093 test results.
RESULTS: Participating clinics submitted an average of 11 DUWL tests per year. Overall, 21% of tests failed, and a moderate positive association (rs =.52, p < 0.001) was found between clinics' DUWL testing frequency and failure rate. Only 7% of failed DUWL tests were followed up by collection of a subsequent test within two weeks, of which 47% still exceeded the 500 CFU/mL threshold.
CONCLUSIONS: Our findings demonstrate an association between DUWL testing frequency and detection of unacceptable microbial levels, along with infrequent retesting and often-inadequate intervention after a failed test. This suggests the need for further efforts at the regulatory and educational levels to maintain adequate water quality during dental treatment.
CLINICAL RELEVANCE: Procedural water can become contaminated in DUWLs and endanger patients. Regular DUWL monitoring and evidence-based interventions to treat contaminated systems are necessary to safeguard patient health.
MATERIALS AND METHODS: Dental clinics (n = 137) aseptically collected samples of output water from their air/water syringes, handpieces, and ultrasonic scaler lines using Sigma-Aldrich® waterline test kits and delivered them to a quality assurance laboratory. Tests were incubated for seven days at room temperature, and those with heterotrophic plate counts > 500 CFU/mL were reported as failures. Statistical analyses were performed on a database containing 4,093 test results.
RESULTS: Participating clinics submitted an average of 11 DUWL tests per year. Overall, 21% of tests failed, and a moderate positive association (rs =.52, p < 0.001) was found between clinics' DUWL testing frequency and failure rate. Only 7% of failed DUWL tests were followed up by collection of a subsequent test within two weeks, of which 47% still exceeded the 500 CFU/mL threshold.
CONCLUSIONS: Our findings demonstrate an association between DUWL testing frequency and detection of unacceptable microbial levels, along with infrequent retesting and often-inadequate intervention after a failed test. This suggests the need for further efforts at the regulatory and educational levels to maintain adequate water quality during dental treatment.
CLINICAL RELEVANCE: Procedural water can become contaminated in DUWLs and endanger patients. Regular DUWL monitoring and evidence-based interventions to treat contaminated systems are necessary to safeguard patient health.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app