Factors affecting electrogram sensing in insertable cardiac monitor: Insights from surface electrocardiogram mapping analysis.

BACKGROUND: Fidelity of electrogram sensing may reduce false alerts from insertable cardiac monitor (ICM).

OBJECTIVE: To assess impact of vector length, implant angle and patient factors on electrogram sensing using surface electrocardiogram (EKG) mapping.

METHODS: Twelve separate precordial single-lead surface EKGs were acquired from 150 participants at two inter-electrode distance (75mm & 45mm), three vector angles (vertical, oblique, horizontal) and in two postures (upright & supine). A subset of fifty patients also received a clinically indicated ICM implant in 1:1 ratio (Medtronic Reveal LINQ:Biotronik Biomonitor III). All EKG and ICM electrogram were analyzed by blinded investigators using DigitizeIt software. P-wave visibility threshold was set at >0.015mV. Logistic regression was used to identify factors impacting P-wave amplitude.

RESULTS: A total of 1,800 tracings from 150 participants [44.5% female, median 59 years old] were assessed. Median P- and R-wave were 45% and 53% larger with vector length of 75 vs. 45mm respectively (both p<0.001). The oblique orientation yielded the best P- and R-wave sensing while posture change did not impact on P-wave amplitude. Mixed effects modeling found that visible P-waves occur more frequently with vector length of 75mm than 45mm (86% vs. 75% respectively, p<.0001). A longer vector length improved both P-wave amplitude and visibility in all BMI categories. There was moderate correlation of P- and R-waves from the ICM electrogram to surface EKG recordings (ICC 0.74 and 0.80, respectively).

CONCLUSION: Longer vector length and oblique implant angle yielded the best electrogram sensing and are relevant considerations for ICM implantations.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app