Add like
Add dislike
Add to saved papers

A Vexing Problem of Air Trapping in the Trima Accel During Plateletpheresis Procedure: Reflecting Our Experience in Troubleshooting.

Curēus 2023 May
Troubleshooting for any fault of apheresis equipment or kit is hardly addressed. Here, we report a unique problem of air trapping in a kit at two different positions leading to failure of the plateletpheresis procedure. Two plateletpheresis procedures were aborted due to "Access pressure low", though the needle position was absolutely perfect. In the third event, platelet yield was not increasing even after 30 minutes from the time of initiation. It was completed after stopping the centrifuge pump which could have displaced the air bubble from the collection port. The root cause for these events was analyzed in consultation with the apheresis technical expert and "air block" was found to be the cause. Air block can also result in a "low access pressure" alarm despite improper phlebotomy being the common cause. Perfect kit loading, checking of tubing defects prior to loading, and comparative analysis of troubleshooting to have adequate knowledge are essential tools for the smooth functioning of apheresis.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app