Agreement between a new optical low coherence reflectometry biometer and an anterior segment optical coherence tomographer.
Eye and Vision (London, England) 2023 Februrary 26
BACKGROUND: To assess agreement of measurements between a new optical low coherence reflectometry (OLCR) biometer (SW-9000, Suoer, Tianjin, China) and a spectral-domain optical coherence tomographer (SD-OCT)/Placido topographer (MS-39, CSO, Florence, Italy) in healthy subjects.
METHODS: A total of 66 right eyes from 66 healthy subjects were enrolled in this prospective study. Three consecutive measurements were randomly obtained with both devices by the same experienced operator to assess agreement. Bland-Altman plots and 95% limits of agreement (LoA) were used to verify the agreement between the devices. Results are presented as mean ± standard deviation (SD).
RESULTS: The SD-OCT/Placido tomographer showed high agreement with the OLCR biometer for all parameters included in this study. The mean differences of central corneal thickness (CCT), anterior chamber depth (ACD), aqueous depth (AQD), mean keratometry (Km) and corneal diameter (CD) were 2.21 ± 2.67 μm (P < 0.001), - 0.10 ± 0.03 mm (P < 0.001), - 0.10 ± 0.04 mm (P < 0.001), - 0.01 ± 0.22 D (P = 0.773) and 0.20 ± 0.16 mm (P < 0.001), respectively. This implies that the inter-device difference in Km was not statistically significant, while the differences in CCT, ACD, AQD, CD were statistically but not clinically significant. The 95% LoAs of CCT, ACD, AQD, Km and CD were - 3.01 to 7.44 μm, - 0.16 to - 0.05 mm, - 0.18 to - 0.03 mm, - 0.45 to 0.43 D, and - 0.12 to 0.51 mm, respectively.
CONCLUSIONS: For CCT, ACD, AQD, Km, and CD in healthy subjects, the new OLCR biometer has high agreement with the SD-OCT/Placido tomographer and can be used interchangeably due to the narrow range of 95% LoAs.
METHODS: A total of 66 right eyes from 66 healthy subjects were enrolled in this prospective study. Three consecutive measurements were randomly obtained with both devices by the same experienced operator to assess agreement. Bland-Altman plots and 95% limits of agreement (LoA) were used to verify the agreement between the devices. Results are presented as mean ± standard deviation (SD).
RESULTS: The SD-OCT/Placido tomographer showed high agreement with the OLCR biometer for all parameters included in this study. The mean differences of central corneal thickness (CCT), anterior chamber depth (ACD), aqueous depth (AQD), mean keratometry (Km) and corneal diameter (CD) were 2.21 ± 2.67 μm (P < 0.001), - 0.10 ± 0.03 mm (P < 0.001), - 0.10 ± 0.04 mm (P < 0.001), - 0.01 ± 0.22 D (P = 0.773) and 0.20 ± 0.16 mm (P < 0.001), respectively. This implies that the inter-device difference in Km was not statistically significant, while the differences in CCT, ACD, AQD, CD were statistically but not clinically significant. The 95% LoAs of CCT, ACD, AQD, Km and CD were - 3.01 to 7.44 μm, - 0.16 to - 0.05 mm, - 0.18 to - 0.03 mm, - 0.45 to 0.43 D, and - 0.12 to 0.51 mm, respectively.
CONCLUSIONS: For CCT, ACD, AQD, Km, and CD in healthy subjects, the new OLCR biometer has high agreement with the SD-OCT/Placido tomographer and can be used interchangeably due to the narrow range of 95% LoAs.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app