Direct Comparison of Peak Bulk Flow Rate of Programmable Intermittent Epidural Bolus and Manual Epidural Bolus Using a Closed-End Multiorifice Catheter: An Experimental Study.

Anesthesia and Analgesia 2022 October 27
BACKGROUND: The programmable intermittent epidural bolus (PIEB) has been popularized as the optimal delivery technique for labor analgesia. Suggested advantages of this method are less local anesthetic consumption, improved maternal satisfaction, potentially shorter duration of labor, and decreased workload requirements for the anesthesia providers. However, a manual bolus is still routinely used for breakthrough pain when the PIEB is underperforming.

METHODS: We conducted a laboratory-based study to quantify the flow through a multiorifice epidural catheter using the PIEB setting on an epidural pump compared to the manual epidural bolus. Four syringe volumes, 3, 5, 10, and 20 mL, were selected for this experiment. The flow in a manual bolus was also studied with and without the presence of an epidural catheter filter. A generalized estimating equation analysis was done to compare data between the groups.

RESULTS: Regardless of the syringe size, there was a several-fold increase in flow when a manual bolus was used compared to a pump-administered dose, with the highest difference in the peak flow rate observed in 3-mL boluses with up to a 12-fold difference, while the difference was, at most, 7-fold in 5-mL and 10-mL boluses. Manual boluses without a filter achieve a mean peak flow rate higher than manual boluses with a filter.

CONCLUSIONS: Our study found that manual boluses produced a higher flow rate compared to the CADD-Solis epidural pump (Smiths Medical). This study also found that the placement of a particulate filter reduces the flow rates generated while bolusing. Bulk flow rate is directly correlated with induced pressure and solution spread. Because higher bolus pressure has been shown to provide a more efficient distribution of local anesthetic and more efficient pain relief, these results may have impactful clinical significance and will pave the way for future studies.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app