PURPOSE: We evaluated a 10.2-cm-wide, minimally elastic, adhesive wrap-based tourniquet (Solo-T or ST) alongside a 3.8-cm-wide windlass-based tourniquet (Combat Application Tourniquet Generation 7, or CAT) to determine if the tension wrap-tightened ST could deliver hemorrhage control equivalent to the windlass-tightened CAT.
METHODS: A cadaver model was used to simulate lower-thigh femoral arterial hemorrhage at "normal" (146 ± 5mmHg) and "elevated" (471 ± 3mmHg) perfusion pressures (mean ± standard error). Three study participants used the ST and CAT to control hemorrhage during 48 timed trials. Arterial occlusion was established by Doppler ultrasound and tourniquet performance was quantified by under-tourniquet pressure cuffs.
RESULTS: Participants achieved 100% (24/24) occlusion success rates and reported similar ease of use for both tourniquets. Occlusion and application times (mean ± standard error) were similar (p > .05) for the ST and CAT under "normal" (occlusion, ST: 25 ± 2 seconds, CAT: 22 ± 2 seconds; application, ST: 27 ± 2 seconds, CAT: 26 ± 2 seconds) and "elevated" (occlusion, ST: 24 ± 7 seconds, CAT: 24 ± 7 seconds; application, ST: 25 ± 7 seconds, CAT: 25 ± 7 seconds) perfusion alike. The ST mean completion pressures (mean ± standard error) were > 40% lower than the CAT under both "normal" perfusion (ST: 110 ± 20mmHg; CAT: 210 ± 30mmHg; p = 0.009) and "elevated" perfusion (ST: 190 ± 50mmHg; CAT: 340 ± 30mmHg; p = 0.03).
CONCLUSION: The adhesive wrap-based ST tourniquet delivered equivalent hemorrhage control performance at significantly lower completion pressures than the CAT.
METHODS: A cadaver model was used to simulate lower-thigh femoral arterial hemorrhage at "normal" (146 ± 5mmHg) and "elevated" (471 ± 3mmHg) perfusion pressures (mean ± standard error). Three study participants used the ST and CAT to control hemorrhage during 48 timed trials. Arterial occlusion was established by Doppler ultrasound and tourniquet performance was quantified by under-tourniquet pressure cuffs.
RESULTS: Participants achieved 100% (24/24) occlusion success rates and reported similar ease of use for both tourniquets. Occlusion and application times (mean ± standard error) were similar (p > .05) for the ST and CAT under "normal" (occlusion, ST: 25 ± 2 seconds, CAT: 22 ± 2 seconds; application, ST: 27 ± 2 seconds, CAT: 26 ± 2 seconds) and "elevated" (occlusion, ST: 24 ± 7 seconds, CAT: 24 ± 7 seconds; application, ST: 25 ± 7 seconds, CAT: 25 ± 7 seconds) perfusion alike. The ST mean completion pressures (mean ± standard error) were > 40% lower than the CAT under both "normal" perfusion (ST: 110 ± 20mmHg; CAT: 210 ± 30mmHg; p = 0.009) and "elevated" perfusion (ST: 190 ± 50mmHg; CAT: 340 ± 30mmHg; p = 0.03).
CONCLUSION: The adhesive wrap-based ST tourniquet delivered equivalent hemorrhage control performance at significantly lower completion pressures than the CAT.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app