Case Reports
Clinical Trial
Add like
Add dislike
Add to saved papers

Strategy to Reduce Hypercapnia in Robot-Assisted Radical Prostatectomy Using Transcutaneous Carbon Dioxide Monitoring: A Prospective Observational Study.

Purpose: Monitoring end-tidal carbon dioxide partial pressure (PET CO2 ) is a noninvasive, continuous method, but its accuracy is reduced by prolonged capnoperitoneum and the steep Trendelenburg position in robot-assisted radical prostatectomy (RARP). Transcutaneous carbon dioxide partial pressure (PTC CO2 ) monitoring, which is not affected by ventilator-perfusion mismatch, has been suggested as a suitable alternative. We compared the agreement of noninvasive measurements with the arterial carbon dioxide partial pressure (PaCO2 ) over a long period of capnoperitoneum, and investigated its sensitivity and predictive power for detecting hypercapnia.

Patients and Methods: The patients who underwent RARP were enrolled in this study prospectively. Intraoperative measurements of PET CO2 , PTC CO2 , and PaCO2 were analyzed. The primary outcome was the agreement of noninvasive monitoring with PaCO2 during prolonged capnoperitoneum. Bias and precision between noninvasive measurements and PaCO2 were assessed using Bland-Altman analysis. The bias and mean absolute difference were compared using a two-tailed Wilcoxon signed-rank test for pairs. The secondary outcome was the sensitivity and predictive power for detecting hypercapnia. To assess this, the Yates corrected chi-square test and the area under the receiver operating characteristic curve were used.

Results: The study analyzed 219 datasets from 46 patients. Compared with PET CO2 , PTC CO2 had lower bias, greater precision, and better agreement with PaCO2 throughout the RARP. The mean absolute difference in PET CO2 and PaCO2 was larger than that of PTC CO2 and PaCO2, and continued to exceed the clinically acceptable range of 5 mmHg after 1 hour of capnoperitoneum. The sensitivity during capnoperitoneum and overall predictive power of PTC CO2 for detecting hypercapnia were significantly higher than those of PET CO2 , suggesting a greater contribution to ventilator adjustment, to treat hypercapnia.

Conclusion: PTC CO2 monitoring measured PaCO2 more accurately than PET CO2 monitoring during RARP requiring prolonged capnoperitoneum and a steep Trendelenburg position. PTC CO2 monitoring also provides more sensitive measurements for ventilator adjustment and detects hypercapnia more effectively than PET CO2 monitoring.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app