We have located links that may give you full text access.
Comparison of hematopoietic progenitor cell collection using different inlet flow rates with the Fenwal Amicus.
Journal of Clinical Apheresis 2022 January 13
PURPOSE: We have used a hematopoietic progenitor cell (HPC) algorithm (standard [STD]) that restricted the inlet flow rate to 65 mL/min for peripheral white blood cell count (PWBC) >35 × 109 /L (STD). In this study, we evaluated a technique that allows 85 mL/min, regardless of the PWBC count (high). For patients with PWBC >35 × 109 /L, a prospective, randomized comparison of the high flow rate vs the STD PWBC-based flow rate (65 mL/min) was performed, comparing CD34+ and lymphocyte yields, collection efficiencies (CE1), mononuclear cells (MNC), and granulocytes, red blood cell (RBC), and platelet content.
METHODS: The Fenwal Amicus version 4.5 with a heparinized ACD-A anticoagulant (AC) delivered at a 26:1 AC ratio was used. Paired comparisons between high and STD techniques were assessed with Wilcoxon signed rank tests, with P < .05 considered significant. Data are summarized as medians.
RESULTS: Forty patient pairs (autologous) were compared. Diagnoses included primarily multiple myeloma (60%) and lymphoma (37.5%). High had significantly higher median average inlet rates (69 vs 55 mL/min), whole blood processed (20 vs 16 L), and cycles (15 vs 14) than STD. There were no significant differences in pre-procedure counts. Collection contents were (high/STD): 306/328 × 106 CD34+ cells, 48/59% CD34+ CE1 (significant), 0.2/0.2 × 109 /kg lymphocytes, 45/57% lymphocyte CE1, 63/59 × 109 WBC, 15/16 × 109 granulocytes, and 1.9/1.7 × 1011 platelets.
CONCLUSIONS: The simpler, standardized high flow technique did not significantly increase or decrease CD34+ cells or lymphocyte yields, but did significantly decrease CD34+ CE1. The effects on cross-cellular content were minimal and not clinically significant.
METHODS: The Fenwal Amicus version 4.5 with a heparinized ACD-A anticoagulant (AC) delivered at a 26:1 AC ratio was used. Paired comparisons between high and STD techniques were assessed with Wilcoxon signed rank tests, with P < .05 considered significant. Data are summarized as medians.
RESULTS: Forty patient pairs (autologous) were compared. Diagnoses included primarily multiple myeloma (60%) and lymphoma (37.5%). High had significantly higher median average inlet rates (69 vs 55 mL/min), whole blood processed (20 vs 16 L), and cycles (15 vs 14) than STD. There were no significant differences in pre-procedure counts. Collection contents were (high/STD): 306/328 × 106 CD34+ cells, 48/59% CD34+ CE1 (significant), 0.2/0.2 × 109 /kg lymphocytes, 45/57% lymphocyte CE1, 63/59 × 109 WBC, 15/16 × 109 granulocytes, and 1.9/1.7 × 1011 platelets.
CONCLUSIONS: The simpler, standardized high flow technique did not significantly increase or decrease CD34+ cells or lymphocyte yields, but did significantly decrease CD34+ CE1. The effects on cross-cellular content were minimal and not clinically significant.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app