Add like
Add dislike
Add to saved papers

A New Solution to a Rare Problem of Implantable Cardiac Device Hypersensitivity.

Curēus 2021 September
For this research, we have considered a case of a man aged 60 years who developed painless fluid accumulation in the implantable cardiac defibrillator (ICD) pocket site. The cardiovascular implantable electronic device initially appeared to be infected, but it was eventually determined that the cause was an allergic reaction, and a novel solution was implemented. For patients with nickel allergies, treatment typically includes avoiding nickel or replacing with gold-plated devices with new leads. Because of the subclavian vein thrombosis, the gold-plated generator was not replaced. ICD generators were covered with antibiotic-coated envelopes while waiting for the replacement. Hypersensitivity to cardiac devices was effectively treated with this technique. By routinely using the envelope, the very rare incidence of device hypersensitivity could further be reduced.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app