Add like
Add dislike
Add to saved papers

Quality comparison between three-dimensional T2-weighted SPACE and two-dimensional T2-weighted turbo spin echo magnetic resonance images for the brachytherapy planning evaluation of prostate and periprostatic anatomy.

Brachytherapy 2020 May 11
PURPOSE: The purpose of this study was to compare an isotropic three-dimensional (3D) T2-weighted sequence sampling perfection with application-optimized contrasts by using flip angle evolution (SPACE) with an axial two-dimensional T2-weighted turbo spin echo (TSE) sequence with regard to overall image quality and the delineation of normal prostate and periprostatic anatomy for low-dose-rate prostate cancer brachytherapy planning evaluation.

METHODS AND MATERIALS: Patients (n = 69) with prostate cancer who had pelvic magnetic resonance imaging (MRI) for low-dose-rate brachytherapy treatment planning were included. Three radiologists independently assessed the visibility of nine anatomic structures on each sequence by using a 5-point scale and overall image quality by using a 4-point scale. The significance of the differences in diagnostic performance was tested with a Wilcoxon signed rank test.

RESULTS: No significant intersequence differences were found for most (7/9) anatomical structures and overall image quality. The mean scores for visibility of anatomical structures on the 3D SPACE and 2D TSE sequences, respectively, were as follows: the zonal anatomy (3.7; 3.9, p = 0.05), prostate capsule (3.9; 4.0, p = 0.08), neurovascular bundle (2.9; 2.9, p = 0.9), rectoprostatic angle (3.8; 3.8, p = 0.35), rectum (4.2; 4.3, p = 0.26), urethra (3.8; 3.9, p = 0.12), urinary bladder (4.6; 4.6, p = 0.61), and overall image quality (2.9; 2.9, p = 0.33). 3D SPACE was superior for delineation of the genitourinary diaphragm (3.8; 3.6, p = 0.003), whereas 2D TSE was superior for delineation of the seminal vesicles (3.5; 4.0, p < 0.0001).

CONCLUSIONS: Anatomic delineation of the prostatic and periprostatic anatomy provided by the 3D SPACE sequence is as robust in quality as that provided by a conventional 2D TSE sequence with superior delineation of the genitourinary diaphragm. For MRI-based brachytherapy treatment planning, the 3D SPACE sequence with subcentimeter isotropic resolution can replace the 2D TSE sequence and be incorporated into standard MRI protocols.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app