Impact of Leadless Pacemaker Therapy on Cardiac and Atrioventricular Valve Function Through 12 Months of Follow-Up.

BACKGROUND: Endocardial pacemaker leads and right ventricular (RV) pacing are well-known causes of tricuspid valve, mitral valve, and cardiac dysfunction. Lead-related adverse consequences can potentially be mitigated by leadless pacemaker (LP) therapy by eliminating the presence of a transvalvular lead. This study assessed the impact of LP placement on cardiac and valvular structure and function.

METHODS: Echocardiographic studies before and 12±1 months after LP implantation were performed between January 2013 and May 2018 at our center and compared with age- and sex-matched controls of dual-chamber transvenous pacemaker recipients.

RESULTS: A total of 53 patients receiving an LP were included, of whom 28 were implanted with a Nanostim and 25 with a Micra LP device. Tricuspid valve regurgitation was graded as being more severe in 23 (43%) patients at 12±1 months compared with baseline ( P<0.001). Compared with an apical position, an RV septal position of the LP was associated with increased tricuspid valve incompetence (odds ratio, 5.20; P=0.03). An increase in mitral valve regurgitation was observed in 38% of patients ( P=0.006). LP implantation resulted in a reduction of RV function, according to a lower tricuspid annular plane systolic excursion ( P=0.003) and RV tricuspid lateral annular systolic velocity ( P=0.02), and a higher RV Tei index ( P=0.04). LP implantation was further associated with a reduction of left ventricular ejection fraction ( P=0.03) and elevated left ventricular Tei index ( P=0.003). The changes in tricuspid valve regurgitation in the LP group were similar to the changes in the dual-chamber transvenous pacemaker control group (43% versus 38%, respectively; P=0.39).

CONCLUSIONS: LP therapy is associated with an increase in tricuspid valve dysfunction through 12 months of follow-up; yet it was comparable to dual-chamber transvenous pacemaker systems. Furthermore, LP therapy seems to adversely impact mitral valve and biventricular function.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app