We have located links that may give you full text access.
Novel Use of GuideLiner with a Low-Profile Balloon for the Retrieval of Disrupted Balloon Catheter.
International Heart Journal 2018 October 26
We report a case of successful percutaneous retrieval of an unexpectedly disrupted balloon catheter using GuideLiner and a low-profile balloon. The procedure and the mechanism of this novel technique were described in detail with ex-vivo testing. This case demonstrated the utility of the combination of GuideLiner and low-profile balloon as a bail-out for intravascular foreign body.