Intradialytic kinetics of middle molecules during hemodialysis and hemodiafiltration.

BACKGROUND: The kinetics of β2-microglobulin during hemodialysis and hemodiafiltration is well described by a two-compartment model where clearance by the dialyzer is from a central compartment volume that approximates plasma volume and a total distribution volume that approximates extracellular fluid volume. The kinetics of middle molecules with molecular weights larger than β2-microglobulin have not been extensively studied.

METHODS: Intradialytic plasma concentrations and overall dialyzer clearances of β2-microglobulin (11.8 kD), myoglobin (16.7 kD) and complement factor D (24.4 kD) were used to estimate three kinetic parameters from a two-compartment model, namely intercompartmental clearance, central compartment volume and total distribution volume, in hemodialysis patients; these data were collected during two clinical trials of medium cut-off dialyzers (with extended middle molecule removal) during hemodialysis and high-flux dialyzers during hemodialysis and hemodiafiltration. In the current exploratory analyses, the kinetic parameters from all dialyzers were combined. Overall dialyzer clearance was evaluated by total mass removed in the dialysate.

RESULTS: In total, 345 sets of kinetic parameters from 35 patients were determined. Intercompartmental clearance and central compartment volume for myoglobin and complement factor D were smaller (P < 0.001) than those for β2-microglobulin. Independent of middle molecule, intercompartmental clearance and central compartment volume were associated with overall dialyzer clearance (P < 0.001), but total distribution volume was not (P = 0.083).

CONCLUSIONS: A two-compartment kinetic model can only describe intradialytic kinetics of middle molecules with molecular weights larger than β2-microglobulin if the central compartment is small and dependent on overall dialyzer clearance.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app