Post marketing study of hemodynamic and hematological noninvasive readings in a blood bank.

Objectives: This validation test was conducted in the Fujisan Blood Bank, Fortaleza, Brazil and evaluated the noninvasive TensorTip MTX (MTX, Cnoga Medical Ltd.) readings of hemoglobin, hematocrit, red blood cells, blood pressure, and heart rate compared to reference lab device readings. Generally, these parameters are measured from venous or capillary blood samples run on a laboratory analyzer or handheld invasive testing devices. Needle sticks are inconvenient to blood donors with relatively high exposure risks. To our vision, noninvasive determination would be of benefit to blood contributors and medical professional teams; it would be fast and painless.

Methods: A total of 334 subjects were included in the Fujisan blood bank validation (65% male, 35% female). Hemoglobin, hematocrit, and red blood cells, as well as blood pressure and heart rate, were measured noninvasively using the MTX device and were compared to venous blood samples run on two laboratory hematology analyzers (Horiba ABX Micros60 and Siemens blood count analyzer), to digital sphygmomanometer (OMRON BP786) and to manual auscultation. The noninvasive measurement with the appropriate virtual arm cuff setting was performed simultaneously with the blood sample extraction of the reference devices measurement.

Results: There was no statistically significant difference ( p  > 0.05, paired, two-tailed t -test) between the average daily hemoglobin, hematocrit, and red blood cells measurements provided by the MTX device and the laboratory hematology analyzer. In addition, there was no significant difference between the daily blood pressure and heart rate results provided by the MTX device and the digital and manual sphygmomanometers. The error calculated between the MTX and the reference device was found to be sufficiently accurate according to the relevant standards.

Conclusion: The MTX accuracy of noninvasive hemoglobin, hematocrit, red blood cells, blood pressure, and heart rate measurements satisfies the industrial standards; therefore, the device enables more accurate, efficient, and effective patient care.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app