Red Cell Alloantibody Screening: Comparative Analysis of Three Different Technologies.
Transfusion Medicine and Hemotherapy 2018 May
Background: The detection of irregular antibody is a critical issue in the management of red blood cell transfusion according to the Type & Screen (T&S) practice. In order to implement the T&S procedure at our blood bank, we compared three different automated analyzers based on column agglutination technique (CAT) or solid phase red cell adherence assay (SPACA) methods.
Methods: Pre-transfusion antibody screening was performed in 986 patients candidate to elective surgery at low risk for red blood cell transfusion. We tested the following kits: the three-cell panel micro-CAT system ID-DiaCell I-II-III (DiaMed), the four-cell panel solid-phase system Capture-R Ready Screen-4 (Immucor), and the four-cell panel micro-CAT system Serascan Diana-4 (Grifols). Positive results were further investigated using corresponding identification panels, and discrepant results were investigated with all the antibody identification systems.
Results: Among 986 samples, we observed 967 concordant negative results (98.1%), 8 concordant positive results (0.8% of cases), and 11 discrepant results (1.1%). Among discrepant samples, an alloantibody could been identified in two patents (anti-M, detected by Serascan Diana-4 and ID-DiaCell I, II, III; anti-Kpa , detected by Capture-R Ready Screen-4 and Serascan Diana-4).
Conclusion: Among the evaluated technologies, the four-cell panel micro-CAT system displayed the highest sensitivity and specificity with an optimal negative predictive value. These features might be relevant to the routine implementation of the T&S transfusion strategy.
Methods: Pre-transfusion antibody screening was performed in 986 patients candidate to elective surgery at low risk for red blood cell transfusion. We tested the following kits: the three-cell panel micro-CAT system ID-DiaCell I-II-III (DiaMed), the four-cell panel solid-phase system Capture-R Ready Screen-4 (Immucor), and the four-cell panel micro-CAT system Serascan Diana-4 (Grifols). Positive results were further investigated using corresponding identification panels, and discrepant results were investigated with all the antibody identification systems.
Results: Among 986 samples, we observed 967 concordant negative results (98.1%), 8 concordant positive results (0.8% of cases), and 11 discrepant results (1.1%). Among discrepant samples, an alloantibody could been identified in two patents (anti-M, detected by Serascan Diana-4 and ID-DiaCell I, II, III; anti-Kpa , detected by Capture-R Ready Screen-4 and Serascan Diana-4).
Conclusion: Among the evaluated technologies, the four-cell panel micro-CAT system displayed the highest sensitivity and specificity with an optimal negative predictive value. These features might be relevant to the routine implementation of the T&S transfusion strategy.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app