Influence and reliability of lower-limb arterial occlusion pressure at different body positions.
PeerJ 2018
Background: Total arterial occlusive pressure (AOP) is used to prescribe pressures for surgery, blood flow restriction exercise (BFRE) and ischemic preconditioning (IPC). AOP is often measured in a supine position; however, the influence of body position on AOP measurement is unknown and may influence level of occlusion in different positions during BFR and IPC. The aim of this study was therefore to investigate the influence of body position on AOP.
Methods: Fifty healthy individuals (age = 29 ± 6 y) underwent AOP measurements on the dominant lower-limb in supine, seated and standing positions in a randomised order. AOP was measured automatically using the Delfi Personalised Tourniquet System device, with each measurement separated by 5 min of rest.
Results: Arterial occlusive pressure was significantly lower in the supine position compared to the seated position (187.00 ± 32.5 vs 204.00 ± 28.5 mmHg, p < 0.001) and standing position (187.00 ± 32.5 vs 241.50 ± 49.3 mmHg, p < 0.001). AOP was significantly higher in the standing position compared to the seated position (241.50 ± 49.3 vs 204.00 ± 28.5 mmHg, p < 0.001).
Discussion: Arterial occlusive pressure measurement is body position dependent, thus for accurate prescription of occlusion pressure during surgery, BFR and IPC, AOP should be measured in the position intended for subsequent application of occlusion.
Methods: Fifty healthy individuals (age = 29 ± 6 y) underwent AOP measurements on the dominant lower-limb in supine, seated and standing positions in a randomised order. AOP was measured automatically using the Delfi Personalised Tourniquet System device, with each measurement separated by 5 min of rest.
Results: Arterial occlusive pressure was significantly lower in the supine position compared to the seated position (187.00 ± 32.5 vs 204.00 ± 28.5 mmHg, p < 0.001) and standing position (187.00 ± 32.5 vs 241.50 ± 49.3 mmHg, p < 0.001). AOP was significantly higher in the standing position compared to the seated position (241.50 ± 49.3 vs 204.00 ± 28.5 mmHg, p < 0.001).
Discussion: Arterial occlusive pressure measurement is body position dependent, thus for accurate prescription of occlusion pressure during surgery, BFR and IPC, AOP should be measured in the position intended for subsequent application of occlusion.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app