BACKGROUND: Measurement uncertainty (MU) is a parameter associated with the result of a measurement that characterizes its dispersion. We report results for estimating MU following the application of a top-down procedure using only proficiency test data to establish uncertainty levels for various analytes.
METHODS: Data were obtained from 142 laboratories participating in the Beijing Center for Clinical Laboratory (BCCL) proficiency testing/external quality assessment (PT/EQA) schemes. The 24-month study included six selected PT shipments to obtain estimates for 50th percentile (median) and 90th percentile MUs and to compare those estimates to usual analytic goals. The number of laboratory participants varied for each trial. The expanded uncertainty (U) was calculated using a cover factor of k=2 for a confidence interval of 95%. All reproducibility, method and laboratory biases came from the PT/EQA data.
RESULTS: The median U (k=2) ranged from 3.2% (plasma sodium, indirect ion selective electrode) to 32.8% (triglycerides, free glycerol blanking) for clinical chemistry analyte means from participants in the same method group. Immunoassay analyte median U results ranged from 11.3% (CA125 tumor marker, Roche) to 33.8% (prostate-specific antigen [PSA], Abbott). The range for median U was 3.5% (red blood cell [RBC], Abx) to 30.3% (fibrinogen [FBG], other) for hematology and coagulation analytes. The MUs for most analytes satisfied quality requirements.
CONCLUSIONS: The use of PT/EQA data, when available, provides an effective means for estimating uncertainties associated with quantitative measurements. Thus, medical laboratories can calculate their own MUs. Proficiency testing organizers can provide participants with an additional MU estimate using only EQA data, which may be updated at the end of each survey.
METHODS: Data were obtained from 142 laboratories participating in the Beijing Center for Clinical Laboratory (BCCL) proficiency testing/external quality assessment (PT/EQA) schemes. The 24-month study included six selected PT shipments to obtain estimates for 50th percentile (median) and 90th percentile MUs and to compare those estimates to usual analytic goals. The number of laboratory participants varied for each trial. The expanded uncertainty (U) was calculated using a cover factor of k=2 for a confidence interval of 95%. All reproducibility, method and laboratory biases came from the PT/EQA data.
RESULTS: The median U (k=2) ranged from 3.2% (plasma sodium, indirect ion selective electrode) to 32.8% (triglycerides, free glycerol blanking) for clinical chemistry analyte means from participants in the same method group. Immunoassay analyte median U results ranged from 11.3% (CA125 tumor marker, Roche) to 33.8% (prostate-specific antigen [PSA], Abbott). The range for median U was 3.5% (red blood cell [RBC], Abx) to 30.3% (fibrinogen [FBG], other) for hematology and coagulation analytes. The MUs for most analytes satisfied quality requirements.
CONCLUSIONS: The use of PT/EQA data, when available, provides an effective means for estimating uncertainties associated with quantitative measurements. Thus, medical laboratories can calculate their own MUs. Proficiency testing organizers can provide participants with an additional MU estimate using only EQA data, which may be updated at the end of each survey.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app