A comparative assessment of inhomogeneity and finite patient dimension effects in 60 Co and 192 Ir high-dose-rate brachytherapy.

Purpose: To perform a comparative study of heterogeneities and finite patient dimension effects in 60 Co and 192 Ir high-dose-rate (HDR) brachytherapy.

Material and methods: Clinically equivalent plans were prepared for 19 cases (8 breast, 5 esophagus, 6 gynecologic) using the Ir2.A85-2 and the Co0.A86 HDR sources, with a TG-43 based treatment planning system (TPS). Phase space files were obtained for the two source designs using MCNP6, and validated through comparison to a single source dosimetry results in the literature. Dose to water, taking into account the patient specific anatomy and materials (Dw,m ), was calculated for all plans using MCNP6, with input files prepared using the BrachyGuide software tool to analyze information from DICOM RT plan exports.

Results: A general TG-43 dose overestimation was observed, except for the lungs, with a greater magnitude for 192 Ir. The distribution of percentage differences between TG-43 and Monte Carlo (MC) in dose volume histogram (DVH) indices for the planning target volume (PTV) presented small median values (about 2%) for both 60 Co and 192 Ir, with a greater dispersion for 192 Ir. Regarding the organs at risk (OARs), median percentage differences for breast V50% were 3% (5%) for 60 Co (192 Ir). Differences in median skin D2cc were found comparable, with a larger dispersion for 192 Ir, and the same applied to the lung D10cc and the aorta D2cc . TG-43 overestimates D2cc for the rectum and the sigmoid, with median differences from MC within 2% and a greater dispersion for 192 Ir. For the bladder, the median of the difference is greater for 60 Co (~2%) than for 192 Ir (~0.75%), demonstrating however a greater dispersion again for 192 Ir.

Conclusions: The magnitude of differences observed between TG-43 based and MC dosimetry and their smaller dispersion relative to 192 Ir, suggest that 60 Co HDR sources are more amenable to the TG-43 assumptions in clinical treatment planning dosimetry.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app