Integrated Assessment of Left Ventricular Electrical Activation and Myocardial Strain Mapping in Heart Failure Patients: A Holistic Diagnostic Approach for Endocardial Cardiac Resynchronization Therapy, Ablation of Ventricular Tachycardia, and Biological Therapy.

OBJECTIVES: This study sought to test the accuracy of strain measurements based on anatomo-electromechanical mapping (AEMM) measurements compared with magnetic resonance imaging (MRI) tagging, to evaluate the diagnostic value of AEMM-based strain measurements in the assessment of myocardial viability, and the additional value of AEMM over peak-to-peak local voltages.

BACKGROUND: The in vivo identification of viable tissue, evaluation of mechanical contraction, and simultaneous left ventricular activation is currently achieved using multiple complementary techniques.

METHODS: In 33 patients, AEMM maps (NOGA XP, Biologic Delivery Systems, Division of Biosense Webster, a Johnson & Johnson Company, Irwindale, California) and MRI images (Siemens 3T, Siemens Healthcare, Erlangen, Germany) were obtained within 1 month. MRI tagging was used to determine circumferential strain (Ecc ) and delayed enhancement to obtain local scar extent (%). Custom software was used to measure Ecc and local area strain (LAS) from the motion field of the AEMM catheter tip.

RESULTS: Intertechnique agreement for Ecc was good (R2  = 0.80), with nonsignificant bias (0.01 strain units) and narrow limits of agreement (-0.03 to 0.06). Scar segments showed lower absolute strain amplitudes compared with nonscar segments: Ecc (median [first to third quartile]: nonscar -0.10 [-0.15 to -0.06] vs. scar -0.04 [-0.06 to -0.02]) and LAS (-0.20 [-0.27 to -0.14] vs. -0.09 [-0.14 to -0.06]). AEMM strains accurately discriminated between scar and nonscar segments, in particular LAS (area under the curve: 0.84, accuracy = 0.76), which was superior to peak-to-peak voltages (nonscar 9.5 [6.5 to 13.3] mV vs. scar 5.6 [3.4 to 8.3] mV; area under the curve: 0.75). Combination of LAS and peak-to-peak voltages resulted in 86% accuracy.

CONCLUSIONS: An integrated AEMM approach can accurately determine local deformation and correlates with the scar extent. This approach has potential immediate application in the diagnosis, delivery of intracardiac therapies, and their intraprocedural evaluation.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app