Add like
Add dislike
Add to saved papers

Influence of Anatomic Changes on the Outcomes of Carotid Siphon Aneurysms After Deployment of Flow-Diverter Stents.

Neurosurgery 2018 December 2
BACKGROUND: Flow-diverter stent (FDS) deployment can cause morphological and hemodynamic changes in the carotid siphon (CS), influencing the occlusion rate of aneurysms in this location.

OBJECTIVE: To evaluate morphological changes to the CS after FDS deployment and their relationship with the rate of occlusion of intracranial aneurysms.

METHODS: A cohort of 183 patients with CS aneurysms were treated by deployment of Pipeline® FDS (Medtronic Inc, Dublin, Ireland). Their CSs were classified as type U, V, C, or S, depending on morphology. The posterior and anterior bend angles were measured on strict lateral cerebral angiogram with digital subtraction before FDS deployment, immediately after deployment, and at 6 mo. Differences between angles were analyzed to identify any correlations with rates of aneurysm occlusion, using the O'Kelly-Marotta classification.

RESULTS: FDS deployment was associated with immediate changes in CS morphology. The mean anterior angle increased from 3.97 ± 25.06° to 22.05 ± 25.18° (P  < .001) and the mean posterior angle increased from 71.98 ± 31.27° to 79.43 ± 31.80° (P  <  .001). Multivariate analysis revealed a progressive, statistically significant increase in frequency of complete (grade D) occlusion at 6-mo follow-up with increasing anterior bend angle (prevalence ratios (PR) = 1.42 for increases between 5.3° and 12°, P  =  .017; PR = 1.56 for increases between 12.1° and 27.6°, P  =  .002; PR = 1.83 for increases >27.6°, P  <  .001, all vs increases <5.3°).

CONCLUSION: FDS deployment induces changes in CS morphology. Specifically, increases in mean anterior angle are associated with better radiological results on 6-mo follow-up digital subtraction angiography.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app