Add like
Add dislike
Add to saved papers

Adult reference intervals for IgG subclasses with Siemens immunonephelometric assays in Chinese population.

OBJECTIVE: To determine the adult reference intervals for the Siemens IgG subclass reagents.

METHODS: 636 blood samples of healthy adults were analyzed to determine the level of IgG subclass using the reagents of Siemens immunonephelometric assay with molecular biology kits.

RESULTS: IgGSc reference intervals were as follows: IgG1 4.45-9.76 g/L, IgG2 2.07-8.57 g/L, IgG3 0.08-0.80 g/L and IgG4 0.05-1.54 g/L. There was an excellent correlation between the total IgG and the sum of the IgG subclasses. No significant gender and age differences were observed.

CONCLUSIONS: Our data provide the missing reference intervals and enable the use of the nephelometric IgG subclass reagents in Chinese. The study can offer reference on clinic diagnose.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app