Intraocular pressure monitoring by rebound tonometry in children with myopia.

BACKGROUND/PURPOSE: Topical atropine treatment is generally accepted to retard the progression of myopia, but it is associated with side effects such as photophobia and elevation of intraocular pressure (IOP). IOP measurements in children are challenging. The traditional applanation tonometry by direct contact with the cornea will require patient's cooperation. The rebound tonometer, using a dynamic electromechanical method for measuring IOP, shows good correlation with traditional tonometry. The purpose of this study is to evaluate the IOP of myopic children under atropine treatment using rebound tonometer and to compare the characteristics between rebound tonometry and applanation tonometry.

METHODS: This study is a prospective study measuring IOP by rebound tonometer in myopic children under regular low-dose atropine treatment. We recruited children with refraction error showing myopia over -0.5 D with 0.15%, 0.3%, or 0.5% atropine eye drops use every night or every other night for myopia control. Children with treatment duration of atropine <1 month were excluded from the study. IOP measurements were performed by applanation tonometer (Tono-Pen XL, Reichert) and rebound tonometer (ICARE). The reliability of rebound tonometer was analyzed with percentage. Comparison of IOP between rebound tonometer and applanation tonometry was presented.

RESULTS: The rebound tonometry was well tolerated by all participants and caused no complaints, discomfort, or adverse events. Totally 42 myopic eyes of 42 subjects were included in the study. The average age of these participants was 10 years old, range from 5 to 16. Median = 10 years old. The average IOP of the right eye by rebound tonometer was 17.4 ± 3 mmHg, and 17.1 ± 3 mmHg by applanation tonometry. Nearly 19%, 33%, and 24% of difference of IOP readings between rebound tonometer and Tono-Pen applanation are within 0 mmHg, 1 mmHg, and 1-2 mmHg, respectively.

CONCLUSIONS: Rebound tonometry has good correlation with applanation tonometry and 76.1% of differences between two tonometers are <2 mmHg. The advantage of drop-free rebound tonometry has made it easier to obtain IOP readings in myopia children under atropine treatment.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app