Clinical Trial
Journal Article
Multicenter Study
Add like
Add dislike
Add to saved papers

Predictors of Paravalvular Regurgitation After Implantation of the Fully Repositionable and Retrievable Lotus Transcatheter Aortic Valve (from the REPRISE II Trial Extended Cohort).

Paravalvular leak (PVL) after transcatheter aortic valve replacement is associated with worse long-term outcomes. The Lotus Valve incorporates an innovative Adaptive Seal designed to minimize PVL. This analysis evaluated the incidence and predictors of PVL after implantation of the Lotus transcatheter aortic valve. The REPRISE II (REpositionable Percutaneous Replacement of Stenotic Aortic Valve through Implantation of Lotus Valve System - Evaluation of Safety and Performance) Study With Extended Cohort enrolled 250 high-surgical risk patients with severe symptomatic aortic stenosis. Aortic regurgitation was assessed by echocardiography pre-procedure, at discharge and 30 days, by an independent core laboratory. Baseline and procedural predictors of mild or greater PVL at 30 days (or at discharge if 30-day data were not available) were determined using a multivariate regression model (n = 229). Of the 229 patients, 197 (86%) had no/trace PVL, 30 had mild, and 2 had moderate PVL; no patient had severe PVL. Significant predictors of mild/moderate PVL included device:annulus area ratio (odds ratio [OR] 0.87; 95% CI 0.83 to 0.92; p <0.001), left ventricular outflow tract calcium volume (OR 2.85; 95% CI 1.44 to 5.63; p = 0.003), and annulus area (OR 0.89; 95% CI 0.82 to 0.96; p = 0.002). When the device:annulus area ratio was <1, the rate of mild/moderate PVL was 53.1% (17 of 32). The rates of mild/moderate PVL with 0% to 5%, 5% to 10%, and >10% annular oversizing by area were 17.5% (11 of 63), 2.9% (2 of 70), and 3.2% (2 of 63), respectively. Significant independent predictors of PVL included device:annulus area ratio and left ventricular outflow tract calcium volume. When the prosthetic valve was oversized by ≥5%, the rate of mild or greater PVL was only 3%. In conclusion, the overall rates of PVL with the Lotus Valve are low and predominantly related to device/annulus areas and calcium; these findings have implications for optimal device sizing.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app