Add like
Add dislike
Add to saved papers

Evaluation of the distribution of activation inside a compact medical cyclotron.

The distribution of activation inside a compact medical cyclotron was evaluated by measuring 1cm dose equivalent rates and γ-ray spectra. Analysis of the distribution of activation showed high activation at the deflector and the magnetic channel. Radionuclides60 Co,57 Co,65 Zn, and54 Mn were detected. Different radionuclides were generated from different components of the cyclotron, and low-activity radionuclides could be detected under low-background-radiation conditions.