Add like
Add dislike
Add to saved papers

Technical and clinical validation of the Greiner FC-Mix glycaemia tube.

BACKGROUND: Measurement of adequate glucose concentrations is complicated by in vitro breakdown of glucose due to glycolysis. Unlike the commonly used NaF-EDTA and NaF-oxalate phlebotomy tubes, citrated NaF-EDTA tubes are reported to directly and thereby completely inhibit glycolysis. Recently, Greiner introduced the Vacuette® FC-Mix NaF-EDTA-citrate tube, currently the only NaF-citrate tube without volume-disturbing liquid additions available on the European market. Here we present its potential as alternative for the laborious and therefore unfeasible conditions for glucose sampling as recommended by the World Health Organization (WHO).

METHODS: The FC-Mix tube was tested against the WHO recommended method of optimal laboratory conditions, both in healthy volunteers and pregnant woman undergoing oral glucose tolerance test (oGTT) for screening of gestational diabetes mellitus (GDM). Glucose concentrations were measured after different incubation times (0-48 h) and temperatures (room temperature, 37 °C), both in uncentrifuged whole blood and centrifuged material.

RESULTS: Deming regression analysis shows that glucose concentrations measured in the FC-Mix tube correlate to the WHO recommended method. Stability is maintained at room temperature for 48 h and at least 24 h at 37 °C. The use of the FC-Mix tube was also validated in screening for GDM and proved comparable to the WHO recommended method in diagnostic outcome.

CONCLUSIONS: The new Greiner FC-Mix tube combines the easy handling of a routine tube with dry additive with the ability to immediately inhibit glycolysis as in the WHO method for optimal pre-analytical and analytical conditions and performs equally to those conditions when screening for GDM.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app