A Comparison of the Cyclic Fatigue Resistance of Used and New Glide Path Files.
Journal of Endodontics 2017 March
INTRODUCTION: The purpose of this study was to compare the cyclic fatigue resistance of used and new ProGlider (Dentsply Maillefer, Ballaigues, Switzerland) and PathFile (Dentsply Maillefer) rotary glide path files.
METHODS: Forty ProGlider and 40 PathFile #2 instruments were used in the present study. In group 1, 20 PathFile #2 files and, in group 2, 20 ProGlider files were used to prepare a total of 40 J-shaped transparent acrylic blocks. In group 3 (control group), 20 new PathFile files and, in group 4 (control group), 20 new ProGlider files were included. Cyclic fatigue testing of instruments was performed in a stainless steel artificial canal with a 5-mm radius of curvature and a 60° angle of curvature. All 80 instruments were rotated until fracture, and the number of cycles to failure (NCF) was recorded. Data were analyzed using 1-way analysis of variance to determine any statistical difference; the significance was determined at the 95% confidence level.
RESULTS: The used glide path files' NCF was lower than the new ones; however, this difference was not statistically significant (P > .05). The cyclic fatigue resistance of the used and new PathFile #2 files was statistically significantly higher than that of the ProGlider files (P < .05).
CONCLUSIONS: Within the limitations of the present study, the used glide path files' NCF was lower than the new ones; however, this difference was not statistically significant.
METHODS: Forty ProGlider and 40 PathFile #2 instruments were used in the present study. In group 1, 20 PathFile #2 files and, in group 2, 20 ProGlider files were used to prepare a total of 40 J-shaped transparent acrylic blocks. In group 3 (control group), 20 new PathFile files and, in group 4 (control group), 20 new ProGlider files were included. Cyclic fatigue testing of instruments was performed in a stainless steel artificial canal with a 5-mm radius of curvature and a 60° angle of curvature. All 80 instruments were rotated until fracture, and the number of cycles to failure (NCF) was recorded. Data were analyzed using 1-way analysis of variance to determine any statistical difference; the significance was determined at the 95% confidence level.
RESULTS: The used glide path files' NCF was lower than the new ones; however, this difference was not statistically significant (P > .05). The cyclic fatigue resistance of the used and new PathFile #2 files was statistically significantly higher than that of the ProGlider files (P < .05).
CONCLUSIONS: Within the limitations of the present study, the used glide path files' NCF was lower than the new ones; however, this difference was not statistically significant.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app