Analytical evaluation of the ADAMS(™) A1c HA 8180 thalassemia mode high-pressure liquid chromatography analyser for the measurement of HbA2 and HbF.
International Journal of Laboratory Hematology 2016 December
BACKGROUND: ADAMS(™) A1cHA-8180T is a HPLC system; within 3.5 min, it quantifies HbF, HbA2 , and HbA0 and flags abnormal peaks. We evaluate its analytical performance for routine estimation of HbA2 and HbF, and critical tests were performed for identifying β-thalassemia carriers.
METHODS: Trueness imprecision, carry over, linearity, and effect of anemia were evaluated according to ICLH, ICLS, or manufacture's guidelines. Comparison (ADAMS(™) A1c HA-8160T) was performed by running 400 samples from healthy subjects, 30 alpha and 80 beta carriers (range: 1.9-5.7 %).
RESULTS: Trueness - HbA2 2.7 %, bias 0.81 %; HbA2 5.8 %, bias 0.38 %. HbA2 4.0% is not affected by Hb in the range 221-40 g/L. Carry over was negligible. Within run: normal control - CV 1.5 %, high control - CV 0.9 %.Within laboratory: normal control - total CV% 1.59%; high control - 0.92 %. Linearity - y = 1.034x - 0.17, R(2 ) = 0.998 (range: 2.8-4.8%).Method comparison - y = 0.93x + 0.22, R(2) = 0.997. HbF imprecision CVs between 0.66 and 1.24% and trueness between 0 and 2.8%. Linearity - y = 1.088x - 0.27, R(2) = 0.999 (0.1-5.7%).
CONCLUSIONS: ADAMS(™) A1c HA-8180T provides a rapid and reliable separation of HbA2 . The measurement is accurate and reproducible, which is needed because of the slight difference between normal and pathological values. The gap in HbA2 values between normal subjects and β-thalassemia carriers makes this an appropriate method for rapid screening for carriers.
METHODS: Trueness imprecision, carry over, linearity, and effect of anemia were evaluated according to ICLH, ICLS, or manufacture's guidelines. Comparison (ADAMS(™) A1c HA-8160T) was performed by running 400 samples from healthy subjects, 30 alpha and 80 beta carriers (range: 1.9-5.7 %).
RESULTS: Trueness - HbA2 2.7 %, bias 0.81 %; HbA2 5.8 %, bias 0.38 %. HbA2 4.0% is not affected by Hb in the range 221-40 g/L. Carry over was negligible. Within run: normal control - CV 1.5 %, high control - CV 0.9 %.Within laboratory: normal control - total CV% 1.59%; high control - 0.92 %. Linearity - y = 1.034x - 0.17, R(2 ) = 0.998 (range: 2.8-4.8%).Method comparison - y = 0.93x + 0.22, R(2) = 0.997. HbF imprecision CVs between 0.66 and 1.24% and trueness between 0 and 2.8%. Linearity - y = 1.088x - 0.27, R(2) = 0.999 (0.1-5.7%).
CONCLUSIONS: ADAMS(™) A1c HA-8180T provides a rapid and reliable separation of HbA2 . The measurement is accurate and reproducible, which is needed because of the slight difference between normal and pathological values. The gap in HbA2 values between normal subjects and β-thalassemia carriers makes this an appropriate method for rapid screening for carriers.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app