Aortic Root Biomechanics After Sleeve and David Sparing Techniques: A Finite Element Analysis.
Annals of Thoracic Surgery 2017 May
BACKGROUND: Aortic root aneurysm can be treated with valve-sparing procedures. The David and Yacoub techniques have shown excellent long-term results but are technically demanding. Recently, a new and simpler procedure, the Sleeve technique, was proposed with encouraging results. We aimed to quantify the biomechanics of the initially aneurysmal aortic root (AR) after the Sleeve procedure to assess whether it induces abnormal stresses, potentially undermining its durability.
METHODS: Two finite element (FE) models of the physiologic and aneurysmal AR were built, accounting for the anatomical asymmetry and the nonlinear and anisotropic mechanical properties of human AR tissues. On the aneurysmal model, the Sleeve and David techniques were simulated based on the corresponding published technical features. Aortic root biomechanics throughout 2 consecutive cardiac cycles were computed in each simulated configuration.
RESULTS: Both sparing techniques restored physiologic-like kinematics of aortic valve (AV) leaflets but induced different leaflets stresses. The time course averaged over the leaflets' bellies was 35% higher in the David model than in the Sleeve model. Commissural stresses, which were equal to 153 and 318 kPa in the physiologic and aneurysmal models, respectively, became 369 and 208 kPa in the David and Sleeve models, respectively.
CONCLUSIONS: No intrinsic structural problems were detected in the Sleeve model that might jeopardize the durability of the procedure. If corroborated by long-term clinical outcomes, the results obtained suggest that using this new technique could successfully simplify the surgical repair of AR aneurysms and reduce intraoperative complications.
METHODS: Two finite element (FE) models of the physiologic and aneurysmal AR were built, accounting for the anatomical asymmetry and the nonlinear and anisotropic mechanical properties of human AR tissues. On the aneurysmal model, the Sleeve and David techniques were simulated based on the corresponding published technical features. Aortic root biomechanics throughout 2 consecutive cardiac cycles were computed in each simulated configuration.
RESULTS: Both sparing techniques restored physiologic-like kinematics of aortic valve (AV) leaflets but induced different leaflets stresses. The time course averaged over the leaflets' bellies was 35% higher in the David model than in the Sleeve model. Commissural stresses, which were equal to 153 and 318 kPa in the physiologic and aneurysmal models, respectively, became 369 and 208 kPa in the David and Sleeve models, respectively.
CONCLUSIONS: No intrinsic structural problems were detected in the Sleeve model that might jeopardize the durability of the procedure. If corroborated by long-term clinical outcomes, the results obtained suggest that using this new technique could successfully simplify the surgical repair of AR aneurysms and reduce intraoperative complications.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app