Contrast-Enhanced High-Pitch Computed Tomography in Pediatric Patients Without Electrocardiography Triggering and Sedation: Comparison of Cardiac Image Quality With Conventional Multidetector Computed Tomography.
Journal of Computer Assisted Tomography 2017 January
OBJECTIVE: The aim of this study was to compare image quality of anatomical cardiac details without electrocardiography triggering, sedation, or heart rate-regulating drugs in contrast-enhanced pediatric chest computed tomography (CT), using high-pitch CT (HPCT) versus conventional pitch-mode technique (multidetector CT [MDCT]).
METHODS: After written informed consent, 55 patients (median age, 11 years; range, 3-17 years) were prospectively included in this institutional review board-approved study. Patients underwent clinically indicated, contrast-enhanced MDCT or HPCT of the chest.Image quality was assessed evaluating morphological criteria on a 3-point scale (from 1, high, to 3, low detail) and summed up in a global score (from 4, best, to 12, poor image quality). Artifacts were analyzed correspondingly (from 3, low, to 9, severe artifacts). Effective dose and size-specific dose estimate were calculated for all scans.
RESULTS: Cardiac image quality was higher in HPCT than in MDCT (7.1 [1.6] vs 8.8 [1.9], P < 0.001). Nevertheless, HPCT showed limitations in image quality, especially concerning the heart valves (2.5 [0.6] and 1.7 [0.5]) and coronary arteries (1.8 [0.6). Artifact score (3.4 [0.6 vs 5.1 [0.9, P < 0.001), effective dose (1.6 [1.3] vs 2.3 [1.6] mSv, P = 0.047), and size-specific dose estimate (2.5 [1.7] vs 4.1 [2.3] mGy, P = 0.002) were lower in HPCT compared with those in MDCT.
CONCLUSIONS: In pediatric patients, contrast-enhanced HPCT of the chest provides high image quality without electrocardiography triggering or sedation, although image quality is somewhat limited for a detailed depiction of cardiac anatomy.
METHODS: After written informed consent, 55 patients (median age, 11 years; range, 3-17 years) were prospectively included in this institutional review board-approved study. Patients underwent clinically indicated, contrast-enhanced MDCT or HPCT of the chest.Image quality was assessed evaluating morphological criteria on a 3-point scale (from 1, high, to 3, low detail) and summed up in a global score (from 4, best, to 12, poor image quality). Artifacts were analyzed correspondingly (from 3, low, to 9, severe artifacts). Effective dose and size-specific dose estimate were calculated for all scans.
RESULTS: Cardiac image quality was higher in HPCT than in MDCT (7.1 [1.6] vs 8.8 [1.9], P < 0.001). Nevertheless, HPCT showed limitations in image quality, especially concerning the heart valves (2.5 [0.6] and 1.7 [0.5]) and coronary arteries (1.8 [0.6). Artifact score (3.4 [0.6 vs 5.1 [0.9, P < 0.001), effective dose (1.6 [1.3] vs 2.3 [1.6] mSv, P = 0.047), and size-specific dose estimate (2.5 [1.7] vs 4.1 [2.3] mGy, P = 0.002) were lower in HPCT compared with those in MDCT.
CONCLUSIONS: In pediatric patients, contrast-enhanced HPCT of the chest provides high image quality without electrocardiography triggering or sedation, although image quality is somewhat limited for a detailed depiction of cardiac anatomy.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app