Study on the measurement of photo-neutron for15 MV photon beam from medical linear accelerator under different irradiation geometries using passive detectors.

AIM OF STUDY: The photo-neutron dose equivalents of 15 MV Elekta precise accelerators were measured for different depths in phantom, for various field sizes, at different distances from the isocenter in the patient plane and for various wedged fields.

MATERIALS AND METHODS: Fast and thermal neutrons are measured using passive detectors such as Columbia Resin-39 and pair of thermoluminescent dosimetry (TLD) 600 and TLD 700 detector from Elekta medical linear accelerator.

RESULTS: It is found that fast photo-neutron dose rate decreases as the depth increases, with a maximum of 0.57 ± 0.08 mSv/Gy photon dose at surface and minimum of 0.09 ± 0.02 mSv/Gy photon dose at 15 cm depth of water equivalent phantom with 10 cm backscatter. Photo neutrons decreases from 1.28 ± 0.03 mSv/Gy to 0.063 ± 0.032 when measured at isocenter and at 100 cm far from the field edge along the longitudinal direction in the patient plane. Fast and thermal neutron doses increases from 0.65 ± 0.05 mSv/Gy to 1.08 ± 0.07 mSv/Gy as the field size increases; from 5 cm × 5 cm to 30 cm × 30 cm for fast neutrons. With increase in wedge field angle from 0° to 60°, it is observed that the fast neutron dose increases from 0.42 ± 0.03 mSv/Gy to 0.95 ± 0.05 mSv/Gy.s

CONCLUSIONS: Measurements indicate the photo-neutrons at few field sizes are slightly higher than the International Electrotechnical Commission standard specifications. Photo-neutrons from Omni wedged fields are studied in details. These studies of the photo-neutron energy response will enlighten the neutron dose to radiation therapy patients and are expected to further improve radiation protection guidelines.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app