Add like
Add dislike
Add to saved papers

Improving Spanning-Knee External Fixator Stiffness: A Biomechanical Study.

The purpose of this study was to test and compare external fixator construct stiffness using pin-to-bar clamps or multipin clamps across 2 external fixation systems. Constructs were tested with 8-mm and 11-mm-diameter bar systems and pin-to-bar or multipin clamps. Three construct designs were tested: construct 1 with a single crossbar and pin-to-bar clamps, construct 2 with 2 crossbars and pin-to-bar clamps, and construct 3 with 2 crossbars and multipin clamps. The stiffness of each construct (N = 24) was tested using anterior-posterior bending. Two crossbars and pin-to-bar clamps resulted in the highest mean stiffness. Constructs with a single crossbar and pin-to-bar clamps had a similar average stiffness compared with constructs with 2 crossbars and multipin clamps. Pin-to-bar clamps with 2 crossbars result in stronger spanning-knee external fixators than constructs using multipin clamps.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app