Pulmonary vein isolation using the Rhythmia mapping system: Verification of intracardiac signals using the Orion mini-basket catheter.
BACKGROUND: During pulmonary vein isolation (PVI), a circular lasso catheter is positioned at the junction between the left atrium (LA) and the pulmonary vein (PV) to confirm PVI. The Rhythmia mapping system uses the Orion mini-basket catheter with 64 electrodes instead of the lasso catheter. However, its feasibility to determine PVI has not been studied.
OBJECTIVE: The purpose of this study was to compare signals between the mini-basket and lasso catheters at the LA-PV junction.
METHODS: In 12 patients undergoing PVI using Rhythmia, the mini-basket and lasso catheters were placed simultaneously at the LA-PV junction for baseline and post-PVI signal assessment. Pacing from both catheters was performed to examine the presence of exit block.
RESULTS: At baseline, recordings of LA and PV potentials were concordant in all PVs. However, after PVI, concordance between the catheters was only 68%. Discordance in all cases resulted from loss of PV potentials on the lasso catheter with persistence of PV potentials on the mini-basket catheter. In 9 of 13 PVs (69%), these potentials represented true PV potentials that were exclusively recorded with the smaller and closely spaced mini-basket electrodes. In the other 4 PVs (31%), these potentials originated from neighboring structures and resulted in underestimation of PVI.
CONCLUSION: The use of the mini-basket catheter alone is sufficient to determine PVI. While it improves recording of PV potentials after incomplete ablation, it is also associated with frequent recording of "PV-like" potentials originating from neighboring structures. In these cases, pacing maneuvers are helpful to determine PVI and avoid excessive ablation.
OBJECTIVE: The purpose of this study was to compare signals between the mini-basket and lasso catheters at the LA-PV junction.
METHODS: In 12 patients undergoing PVI using Rhythmia, the mini-basket and lasso catheters were placed simultaneously at the LA-PV junction for baseline and post-PVI signal assessment. Pacing from both catheters was performed to examine the presence of exit block.
RESULTS: At baseline, recordings of LA and PV potentials were concordant in all PVs. However, after PVI, concordance between the catheters was only 68%. Discordance in all cases resulted from loss of PV potentials on the lasso catheter with persistence of PV potentials on the mini-basket catheter. In 9 of 13 PVs (69%), these potentials represented true PV potentials that were exclusively recorded with the smaller and closely spaced mini-basket electrodes. In the other 4 PVs (31%), these potentials originated from neighboring structures and resulted in underestimation of PVI.
CONCLUSION: The use of the mini-basket catheter alone is sufficient to determine PVI. While it improves recording of PV potentials after incomplete ablation, it is also associated with frequent recording of "PV-like" potentials originating from neighboring structures. In these cases, pacing maneuvers are helpful to determine PVI and avoid excessive ablation.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app