Anterior interbody fusion of the cervical spine with Zero-P spacer: prospective comparative study-clinical and radiological results at a minimum 2 years after surgery.

Spine 2013 June 2
STUDY DESIGN: A prospective study.

OBJECTIVE: The aim of this study was to compare clinical and radiological efficacy of anterior cervical microdiscectomy and fusion done by the newly designed low-profile interbody spacer in cases of symptomatic cervical spine spondylosis.

SUMMARY OF BACKGROUND DATA: There are basically 2 ways to provide interbody fusion in the degenerative cervical spine; the first is by way of an unanchored "stand-alone" bone graft or cage, and the second is with bone graft or a cage anchored with a plate. Both concepts have their own benefits as well as potential drawbacks. Low-profile angle-stable spacer Zero-P is an implant that can potentially limit the drawbacks of both these procedures. METHODS.: Prospective study collecting clinical and radiological data of 77 patients undergoing anterior cervical interbody fusion of 1 or 2 motion segments from C3-C7 was performed. Zero-P spacer was used in 44 patients (55 segments) and in 33 cases (41 segments), stabilization was done using interbody spacer and dynamic anterior cervical plate. Patients were followed a minimum of 2 years after surgery.

RESULTS: There was no significant difference in neck disability index values, presence of dysphagia (P = 0.308), and Cobb C values during follow-up (P = 0.051) between both groups. A significant difference in the first 2 values of Cobb S was found (P < 0.001), but the next course of Cobb S changes showed no difference in either group. No difference was found in the radiological stability during follow-up, and no revision surgery was done.

CONCLUSION: The results of this study confirm biomechanical assumptions associated with the Zero-P spacer. Implantation of this new cage results in setting required biomechanical conditions in the treated segment that are comparable with those when the segment is treated with a dynamic plate. However, the potential of the mentioned implant to reduce the incidence of postoperative dysphagia was not proven on this sample of patients.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app