The case of the floating gel.

In patients with chronic renal failure, blood samples for laboratory analysis are often taken via dialysis catheters. This report describes a case of gross spurious hypernatraemia in a blood sample collected from a patient undergoing haemodialysis. After centrifugation of the blood sample in question, the separator gel formed the topmost layer, with the serum in the middle and the clot at the bottom. Subsequent analysis of the serum showed severe hypernatraemia (serum sodium, 744 mmol/litre). It was established that the blood sample had been taken from the patient's dialysis catheter into which 3 ml of Citra-Lock (46.7% trisodium citrate) had been instilled previously as a "catheter locking" solution. The hypernatraemia seen in this case was recognised immediately as an artefact, but it was found that even minimal contamination of blood samples with Citra-Lock may significantly affect sodium concentrations. This contamination may be missed, with potentially adverse consequences for patient management.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app