JOURNAL ARTICLE
RESEARCH SUPPORT, NON-U.S. GOV'T
Add like
Add dislike
Add to saved papers

Effect of a dual-lumen peripheral catheter on the delivery of known incompatible medications.

OBJECTIVE: To determine the degree to which a dual-lumen peripheral catheter prevented precipitation of solutions known to be incompatible due to pH during simultaneous infusion in an in vitro model.

METHODS: An in vitro model was devised to simulate peripheral venous blood flow from an antecubital source to systemic circulation. Ondansetron was simultaneously infused with fluorouracil, aminophylline, sodium bicarbonate, and ampicillin sodium in concentrations reflective of clinical conditions into the Twin Cath 20/22 (the dual-lumen catheter used in this experiment). Study solutions were primed with the prepared drug solution and administered for 15 minutes. Phase I used Normosol-R as the diluent to gather preliminary data; phase II used human plasma. All samples were obtained immediately before the start of the infusion and at 5, 10, and 15 minutes during the infusion, and 5 minutes after the infusion. Samples were visually inspected at each time point for precipitation and analyzed in duplicate by the appropriate stability-indicating HPLC method (except for sodium bicarbonate). Compatibility was defined as no visual evidence of precipitation and no more than 15% mean change in final versus initial concentration.

RESULTS: Phase I experiments showed immediate precipitation in Normosol-R within the venous flow. However, in phase II, because of the buffering capacity that plasma proteins add to plasma, no precipitation occurred. All the drug combinations used in this study have been reported to be incompatible at the concentrations tested; however, we detected no incompatibilities.

CONCLUSIONS: The results of this study indicate that using a dual-lumen peripheral catheter, such as the Twin Cath, may allow solutions incompatible due to pH to be administered simultaneously.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app