[Blood flow quantification in hemodialysis shunts by phase contrast magnetic resonance angiography (PC-MRA) compared with duplex sonography].
PURPOSE: Aim of this study was to evaluate whether phase contrast MR angiography (PC-MRA) could provide additional functional information besides morphology in the assessment of haemodialysis fistulae.
MATERIAL AND METHOD: Twenty-two patients (11 male, 11 female), aged 22-77 years, were examined. MR images were obtained with a 1.5 T Gyroscan ACS-NT (Philips, Best, Netherlands) using a high-resolution wrap-around coil. In addition to MRA blood-flow measurements were performed with a gradient-echo sequence (TR 14 ms, TE 5-5.5 ms, flip-angle 15 degrees, 6 mm slice thickness, retrospective gating, matrix 96:128) in the venous and arterial section of the fistulae. Doppler flow measurements were performed at the same position with a Sonoline Elegra (Siemens AG, Erlangen) using a 7.5 MHz transducer.
RESULTS: Both methods of flow-volume measurements showed a good correlation (r = 0.94 in the arterial section, r = 0.90 in the venous section, p < 0.001). The average calculated blood flow was measured 11% (arterial section) and 12.8% (venous section) higher with Pulsed Waved Doppler as compared to PC-MRA.
CONCLUSION: PC-MRA with a high-resolution wrap-around coil is a reliable method for measuring functional parameters like flow-volume and flow velocity in haemodialysis fistulae and a useful complement to the visualisation capabilities of MRA.
MATERIAL AND METHOD: Twenty-two patients (11 male, 11 female), aged 22-77 years, were examined. MR images were obtained with a 1.5 T Gyroscan ACS-NT (Philips, Best, Netherlands) using a high-resolution wrap-around coil. In addition to MRA blood-flow measurements were performed with a gradient-echo sequence (TR 14 ms, TE 5-5.5 ms, flip-angle 15 degrees, 6 mm slice thickness, retrospective gating, matrix 96:128) in the venous and arterial section of the fistulae. Doppler flow measurements were performed at the same position with a Sonoline Elegra (Siemens AG, Erlangen) using a 7.5 MHz transducer.
RESULTS: Both methods of flow-volume measurements showed a good correlation (r = 0.94 in the arterial section, r = 0.90 in the venous section, p < 0.001). The average calculated blood flow was measured 11% (arterial section) and 12.8% (venous section) higher with Pulsed Waved Doppler as compared to PC-MRA.
CONCLUSION: PC-MRA with a high-resolution wrap-around coil is a reliable method for measuring functional parameters like flow-volume and flow velocity in haemodialysis fistulae and a useful complement to the visualisation capabilities of MRA.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app