Automated external defibrillation: laboratory evaluation.

Twelve samples of ventricular fibrillation were fed into nine automated external defibrillator-pacemakers ( AEDP , "Heart Aid") of recent design. All the devices recognised and defibrillated ventricular fibrillation in seven of the samples within 30 sec. None of the devices reacted to two of the samples; in the remaining three there was inter-device variation ranging from an appropriate response to no response, as well as inappropriate pacing or delay in recognition and treatment. Poor recognition of some ventricular fibrillation waveforms with considerable inter-device variation limits the usefulness of this model. A new prototype responded more consistently and future models may be of value in community resuscitation. The difficulty of evaluating the diagnostic capability of AEDP devices in clinical use makes comprehensive laboratory testing essential prior to release.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app