BACKGROUND AND PURPOSE: Concerns have emerged regarding elevated levels of cobalt and chromium in patients with metal-on-metal megaprostheses. This prospective study aims to identify systemic cobalt and chromium levels in metal-on-polyethylene knee and hip megaprostheses and their associations with other factors.
METHODS: 56 patients underwent knee or hip megaprosthesis surgery at 2 sarcoma centers. Serum cobalt and chromium levels were measured preoperatively and thrice within the first year using inductively coupled plasma mass spectrometry.
RESULTS: A statistically significant difference in serum cobalt levels (1.4 ppb; 95% confidence interval [CI] 0.0-3.3) was observed 1 year after knee megaprosthesis surgery compared with preoperative levels. In contrast no difference in chromium levels was observed after 1 year compared with preoperative levels (0.05 ppb; CI 0.0-0.8). An association between younger age, higher eGFR, and increased cobalt levels was observed. No significant correlations were found between ion levels and resection length or the number of modular connections.
CONCLUSION: We found elevated serum ion levels in metal-on-polyethylene knee megaprostheses in contrast to metal-on-polyethylene hip megaprostheses. Furthermore, a positive correlation between cobalt and chromium levels, and between cobalt and eGFR was identified, along with a negative correlation between cobalt and age. This study highlights the importance of monitoring systemic cobalt and chromium levels in patients with megaprostheses.
METHODS: 56 patients underwent knee or hip megaprosthesis surgery at 2 sarcoma centers. Serum cobalt and chromium levels were measured preoperatively and thrice within the first year using inductively coupled plasma mass spectrometry.
RESULTS: A statistically significant difference in serum cobalt levels (1.4 ppb; 95% confidence interval [CI] 0.0-3.3) was observed 1 year after knee megaprosthesis surgery compared with preoperative levels. In contrast no difference in chromium levels was observed after 1 year compared with preoperative levels (0.05 ppb; CI 0.0-0.8). An association between younger age, higher eGFR, and increased cobalt levels was observed. No significant correlations were found between ion levels and resection length or the number of modular connections.
CONCLUSION: We found elevated serum ion levels in metal-on-polyethylene knee megaprostheses in contrast to metal-on-polyethylene hip megaprostheses. Furthermore, a positive correlation between cobalt and chromium levels, and between cobalt and eGFR was identified, along with a negative correlation between cobalt and age. This study highlights the importance of monitoring systemic cobalt and chromium levels in patients with megaprostheses.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app