Large Animal Translational Validation of 3 Mitral Valve Repair Operations for Mitral Regurgitation Using a Mitral Valve Prolapse Model: A Comprehensive In Vivo Biomechanical Engineering Analysis.

BACKGROUND: Various mitral repair techniques have been described. Though these repair techniques can be highly effective when performed correctly in suitable patients, limited quantitative biomechanical data are available. Validation and thorough biomechanical evaluation of these repair techniques from translational large animal in vivo studies in a standardized, translatable fashion are lacking. We sought to evaluate and validate biomechanical differences among different mitral repair techniques and further optimize repair operations using a large animal mitral valve prolapse model.

METHODS: Male Dorset sheep (n=20) had P2 chordae severed to create the mitral valve prolapse model. Fiber Bragg grating force sensors were implanted to measure chordal forces. Ten sheep underwent 3 randomized, paired mitral valve repair operations: neochord repair, nonresectional leaflet remodeling, and triangular resection. The other 10 sheep underwent neochord repair with 2, 4, and 6 neochordae. Data were collected at baseline, mitral valve prolapse, and after each repair.

RESULTS: All mitral repair techniques successfully eliminated regurgitation. Compared with mitral valve prolapse (0.54±0.18 N), repair using neochord (0.37±0.20 N; P =0.02) and remodeling techniques (0.30±0.15 N; P =0.001) reduced secondary chordae peak force. Neochord repair further decreased primary chordae peak force (0.21±0.14 N) to baseline levels (0.20±0.17 N; P =0.83), and was associated with lower primary chordae peak force compared with the remodeling (0.34±0.18 N; P =0.02) and triangular resectional techniques (0.36±0.27 N; P =0.03). Specifically, repair using 2 neochordae resulted in higher peak primary chordal forces (0.28±0.21 N) compared with those using 4 (0.22±0.16 N; P =0.02) or 6 neochordae (0.19±0.16 N; P =0.002). No difference in peak primary chordal forces was observed between 4 and 6 neochordae ( P =0.05). Peak forces on the neochordae were the lowest using 6 neochordae (0.09±0.11 N) compared with those of 4 neochordae (0.15±0.14 N; P =0.01) and 2 neochordae (0.29±0.18 N; P =0.001).

CONCLUSIONS: Significant biomechanical differences were observed underlying different mitral repair techniques in a translational large animal model. Neochord repair was associated with the lowest primary chordae peak force compared to the remodeling and triangular resectional techniques. Additionally, neochord repair using at least 4 neochordae was associated with lower chordal forces on the primary chordae and the neochordae. This study provided key insights about mitral valve repair optimization and may further improve repair durability.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app