Numerical optimization of longitudinal collimator geometry for novel x-ray field.

A novel X-ray field produced by an ultrathin conical target is described in the literature. However, the optimal design for an associated collimator remains ambiguous. Current optimization methods using Monte Carlo calculations restrict the efficiency and robustness of the design process. A more generic optimization method that reduces parameter constraints while minimizing computational load is necessary. A numerical method for optimizing the longitudinal collimator hole geometry for a cylindrically-symmetrical X-ray tube is demonstrated and compared to Monte Carlo calculations. 
Approach: The X-ray phase space was modelled as a four-dimensional histogram differential in photon initial position, final position, and photon energy. The collimator was modelled as a stack of thin washers with varying inner radii. Simulated annealing was employed to optimize this set of inner radii according to various objective functions calculated on the photon flux at a specified plane. 
Main results: The analytical transport model used for optimization was validated against Monte Carlo calculations using Geant4 via its wrapper, TOPAS. Optimized collimators and the resulting photon flux profiles are presented for three focal spot sizes and five positions of the source. Optimizations were performed with multiple objective functions based on various weightings of precision, intensity, and field flatness metrics. Finally, a select set of these optimized collimators, plus a parallel-hole collimator for comparison, were modelled in TOPAS. The evolution of the radiation field profiles are presented for various positions of the source for each collimator.
Significance: This novel optimization strategy proved consistent and robust across the range of X-ray tube settings regardless of the optimization starting point. Common collimator geometries were re-derived using this algorithm while simultaneously optimizing geometry-specific parameters. The advantages of this strategy over iterative Monte Carlo-based techniques, including computational efficiency, radiation source-specificity, and solution flexibility, make it a desirable optimization method for complex irradiation geometries. &#xD.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app