Accuracy of Pocket-sized Ultrasound Devices to Evaluate Inferior Vena Cava Diameter and Variability in Critically Ill Patients.
Indian Journal of Critical Care Medicine 2024 April
PURPOSE: By using inferior vena cava (IVC) measurements, clinicians can detect fluid status and responsiveness and find out the etiology of hypotension, acute heart failure, and sepsis easier. Pocket-sized ultrasound devices (PSUD) may take this advantage a few steps further by their lower costs, user-friendly interface, and easily applicable structure.In this study, we aimed to determine the diagnostic value of a PSUD compared with a standard ultrasound device (SD) for the measurement of IVC diameter (IVCD) and its respiratory variability.
MATERIALS AND METHODS: We measured the inspiratory, expiratory diameters of IVC, and calculated the inferior vena cava collapsibility index (IVCCI). We investigated 42 intensive care unit (ICU) patients.
RESULTS: There was no difference in inspiratory (PSUD: 1.34 ± 0.67 cm; SD: 1.35 ± 0.68 cm) and expiratory (PSUD: 1.98 ± 0.53 cm; SD: 2.01 ± 0.49 cm) IVCD among measurements with PSUD and SD ( p > 0.05). There was also no difference between IVCCI's measured with PSUD (39 ± 20%) and SD (39 ± 20%) ( p > 0.05). The Bland-Altman analysis revealed that the width of 95% limits of agreement were similar for both devices. There was a good inter-device agreement among PSUD and SD for measurements of IVCD, and there was no difference between IVCCI's measured using both ultrasound devices.
CONCLUSION: We support that the idea of a PSUD is as reliable as a SD for IVC measurements.
HOW TO CITE THIS ARTICLE: Inci K, Gürsel G. Accuracy of Pocket-sized Ultrasound Devices to Evaluate Inferior Vena Cava Diameter and Variability in Critically Ill Patients. Indian J Crit Care Med 2024;28(4):369-374.
MATERIALS AND METHODS: We measured the inspiratory, expiratory diameters of IVC, and calculated the inferior vena cava collapsibility index (IVCCI). We investigated 42 intensive care unit (ICU) patients.
RESULTS: There was no difference in inspiratory (PSUD: 1.34 ± 0.67 cm; SD: 1.35 ± 0.68 cm) and expiratory (PSUD: 1.98 ± 0.53 cm; SD: 2.01 ± 0.49 cm) IVCD among measurements with PSUD and SD ( p > 0.05). There was also no difference between IVCCI's measured with PSUD (39 ± 20%) and SD (39 ± 20%) ( p > 0.05). The Bland-Altman analysis revealed that the width of 95% limits of agreement were similar for both devices. There was a good inter-device agreement among PSUD and SD for measurements of IVCD, and there was no difference between IVCCI's measured using both ultrasound devices.
CONCLUSION: We support that the idea of a PSUD is as reliable as a SD for IVC measurements.
HOW TO CITE THIS ARTICLE: Inci K, Gürsel G. Accuracy of Pocket-sized Ultrasound Devices to Evaluate Inferior Vena Cava Diameter and Variability in Critically Ill Patients. Indian J Crit Care Med 2024;28(4):369-374.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app