Automated detection of vertebral body misalignments in orthogonal kV and MV guided radiotherapy: application to a comprehensive retrospective dataset.

OBJECTIVE: In image-guided radiotherapy (IGRT), off-by-one vertebral body misalignments are rare but potentially catastrophic. In this study, an automated method to detect such misalignments with IGRT radiographs was investigated using densely-connected convolutional networks (DenseNets) for applications towards real-time error prevention and retrospective error auditing.

APPROACH: A total of 4213 images acquired from 527 radiotherapy patients aligned with planar kV or MV radiographs were used to develop and test error-detection software modules. Digitally reconstructed radiographs (DRRs) and setup images were retrieved and co-registered according to the clinically applied alignment contained in the DICOM REG files. A semi-automated algorithm was developed to simulate patient positioning errors on the anterior-posterior (AP) and lateral (LAT) images shifted by one vertebral body. A DenseNet architecture was designed to classify either AP images individually or AP and LAT image pairs. Receiver-operator characteristic curves (ROC) and areas under the curves (AUC) were computed to evaluate the classifiers on test subsets. Subsequently, the algorithm was applied to the entire dataset in order to retrospectively determine the absolute off-by-one vertebral body error rate for planar radiograph guided RT at our institution from 2011-2021.

MAIN RESULTS: The AUCs for the kV models were 0.98 for unpaired AP and 0.99 for paired AP-LAT. The AUC for the MV AP model was 0.92. For a specificity of 95%, the paired kV model achieved a sensitivity of 99%. Application of the model to the entire dataset yielded a per-fraction off-by-one vertebral body error rate of 0.044% [0.0022%, 0.21%] for paired kV IGRT including one previously unreported error.

SIGNIFICANCE: Our error detection algorithm was successful in classifying vertebral body positioning errors with sufficient accuracy for retrospective quality control and real-time error prevention. The reported positioning error rate for planar radiograph IGRT is unique in being determined independently of an error reporting system.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app