Add like
Add dislike
Add to saved papers

Investigation on the accuracy of field widths and the quantitative relationship with energy variations for dynamic jaws helical tomotherapy.

BACKGROUND AND PURPOSE: TomoEDGE is an advanced technology for TomoTherapy treatment delivery by introducing a sliding-window dynamic jaw motion. The front and back jaws move independently at the start and end of a target volume along the longitudinal couch direction to reduce the undesired dose to the normal tissues. The accuracy of field width is essential to treatment delivery in this regard. The purpose of this work was to analyze the performance of dynamic jaws on helical tomotherapy and investigate the relationship with energy variation.

METHODS: The Tomotherapy-Quality-Assurance (TQA) Dynamic Field Width procedure was performed monthly across three tomotherapy machines. All field widths were analyzed, especially the FWHM of the 10 mm field width. Field width measurements were compared with the ratio of Percentage Depth Dose at 20 and 10 cm to render the value of correlation. Changes in beam FWHM and energy were further discussed. Two-year data were collected for this purpose.

RESULTS: On average, measured field widths in each unit agreed within 1% tolerance recommendation stated. The average absolute difference between reference and measured FWs in each unit was approximately 0.07 mm. An increase of 1.5% in the FW of the 10 mm nominal beam width was correlated with a 1% increase in PDD20,10 ratio, implying a positive correlation between the two factors (p < 0.002).

CONCLUSIONS: A positive correlation between nominal 10 mm FW and PDD20,10 was observed. In the case that the PDD20,10 marginally passes the QA tests, users are recommended to consider further verification on Dynamic Jaws to ensure the smallest field width to be within tolerance, which is essential to maintain effective treatment in TomoEDGE system. Since the regression of this study was a single-factor model, other confounding factors such as the focal spot size of linear accelerator should also be considered when evaluating the machine status.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app