A reproducibility study of knee cartilage volume and thickness values derived by fully automatic segmentation based on three-dimensional dual-echo in steady state data from 1.5 T and 3 T magnetic resonance imaging.

Magma 2023 October 11
OBJECTIVE: To evaluate the repeatability of cartilage volume and thickness values at 1.5 T MRI using a fully automatic cartilage segmentation method and reproducibility of the method between 1.5 T and 3 T data.

METHODS: The study included 20 knee joints from 10 healthy subjects with each subject having undergone double-knee MRI. All knees were scanned at 1.5 T and 3 T MR scanners using a three-dimensional (3D) high-resolution dual-echo in steady state (DESS) sequence. Cartilage volume and thickness of 21 subregions were quantified using a fully automatic cartilage segmentation research application (MR Chondral Health, version 3.0, Siemens Healthcare, Erlangen, Germany). The volume and thickness values derived from fully automatically computed segmentation masks were analyzed for the scan-rescan data from the same volunteers. The accuracy of the automatic segmentation of the cartilage in 1.5 T images was evaluated by the dice similarity coefficient (DSC) and Hausdorff distance (HD) using the manually corrected segmentation as a reference. The volume and thickness values calculated from 1.5 T and 3 T were also compared.

RESULTS: No statistically significant differences were found for cartilage thickness or volume across all subregions between the scan-rescanned data at 1.5 T (P > 0.05). The mean DSC between the fully automatic and manually corrected knee cartilage segmentation contours at 1.5 T was 0.9946. The average value of HD was 2.41 mm. Overall, there was no statistically significant difference in the cartilage volume or thickness in most-subregions between the two field strengths (P > 0.05) except for the medial region of femur and tibia. Bland-Altman plot and intraclass correlation coefficient (ICC) showed high consistency between results obtained based on same and different scanning sequences.

CONCLUSION: The cartilage segmentation software had high repeatability for DESS images obtained from the same device. In addition, the overall reproducibility of the images obtained from equipment of two different field strengths was satisfactory.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app